EasyManuals Logo

Keithley 6517B Reference Manual

Keithley 6517B
498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #404 background imageLoading...
Page #404 background image
Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-153
:ASYNchronous commands
:ILINe <NRf>
This command is used to select the input trigger line arm, scan, and measure layers.
Usage
:ARM[:SEQuence[1]][:LAYer[1]]:TCONfigure:ASYNchronous:ILINe <NRf>
:ARM[:SEQuence[1]]:LAYer2:TCONfigure:ASYNchronous:ILINe <NRf>
:TRIGger[:SEQuence[1]]:TCONfigure:ASYNchronous:ILINe <NRf>
:ILINe?
<NRf>
1: Line #1
2: Line #2
3: Line #3
4: Line #4
5: Line #5
6: Line #6
Details
These commands are used to select an input line for the asynchronous trigger link of the specified layer. Keep in
mind that asynchronous trigger link input and output cannot share the same line. If you assign the input to a line
that is already being used for the output, an error occurs and the command is ignored.
Also see
:OLINe <NRf> (on page 14-153)
:OLINe <NRf>
This command is used to select the output trigger line arm, scan, and measure layers.
Usage
:ARM[:SEQuence[1]][:LAYer[1]]:TCONfigure:ASYNchronous:OLINe <NRf>
:ARM[:SEQuence[1]]:LAYer2:TCONfigure:ASYNchronous:OLINe <NRf>
:TRIGger[:SEQuence[1]]:TCONfigure:ASYNchronous:OLINe <NRf>
:OLINe?
<NRf>
1: Line #1
2: Line #2
3: Line #3
4: Line #4
5: Line #5
6: Line #6
Details
These commands are used to select an output line for the asynchronous trigger link of the specified
layer. Keep in mind that asynchronous trigger link input and output cannot share the same line. If you
assign the output to a line that is already being used for the input, an error occurs and the command
is ignored.
Also see
None

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
Input Impedance>200 TΩ
Electrometer TypeElectrometer/High Resistance Meter
Current Measurement Range20 fA to 20 mA
DisplayVacuum Fluorescent Display
Communication InterfacesIEEE-488 (GPIB), RS-232
Power Supply100 V / 120 V / 220 V / 240 V; 50 Hz / 60 Hz
Resistance Measurement RangeUp to 200 TΩ

Related product manuals