Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-51
[:DATA] <type>[,length]
This command specifies the data format
Usage
:FORMat[:DATA] <type>[,<length>]
[:DATA]?
<type>
ï‚· ASCii: ASCII format
ï‚· REAL: IEEE754 single precision format
ï‚· REAL: IEEE754 double precision format
ï‚· SREal: IEEE754 single precision format
ï‚· DREal: IEEE754 double precision format
<length>
32 or 64; optional for REAL; default 32, single precision format
Details
When using the ASCII, SREal, or DREal parameter, a <length> parameter is not required.
This command is used to select the data format for transferring readings over the bus. For every
reading conversion, the data string sent over the bus contains the elements specified by the
:ELEMents command. The specified elements are sent in a particular order.
The ASCII data format is in a direct readable form for the operator. Most BASIC languages easily
convert ASCII mantissa and exponent to other formats. However, some speed is compromised to
accommodate the conversion. The figure below shows the ASCII format that includes all the data
elements.
Figure 113: ASCII data format

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals