EasyManua.ls Logo

Keithley 6517B

Keithley 6517B
498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 14: SCPI command reference Model 6517B Electrometer Reference Manual
14-38 6517B-901-01 Rev. C / August 2015
:IMMediate
This command is used to perform CALC 2.
Usage
:CALCulate2:IMMediate
:IMMediate?
Details
The :IMMediate command is used to perform the selected CALC 2 operation on the readings in the
buffer (assuming CALC 2 is enabled). After performing the calculation, the result can be read by
using the :CALCulate2:DATA? query command.
An alternate way to perform the calculation and read the result is by using the query form of the
command (:IMMediate?). When this command is sent, the calculation is performed and the result is
queried.
Also see
None
DATA?
This command is used to read the CALC 2 result.
Usage
:CALCulate2:DATA?
Details
This query command is used to read the result of the CALC 2 operation. If CALC 2 is disabled or
NONE is selected, the "raw" reading is read instead.
Note that this command does not initiate a calculation. This command is used only to read the result
of the last calculation.
Also see
None
CALCulate3
This Calculate subsystem is used to configure and control the limit tests (LIMIT 1 test and LIMIT 2 test).

Table of Contents

Other manuals for Keithley 6517B

Related product manuals