Section 14: SCPI command reference Model 6517B Electrometer Reference Manual
14-38 6517B-901-01 Rev. C / August 2015
:IMMediate
This command is used to perform CALC 2.
Usage
:CALCulate2:IMMediate
:IMMediate?
Details
The :IMMediate command is used to perform the selected CALC 2 operation on the readings in the
buffer (assuming CALC 2 is enabled). After performing the calculation, the result can be read by
using the :CALCulate2:DATA? query command.
An alternate way to perform the calculation and read the result is by using the query form of the
command (:IMMediate?). When this command is sent, the calculation is performed and the result is
queried.
Also see
None
DATA?
This command is used to read the CALC 2 result.
Usage
:CALCulate2:DATA?
Details
This query command is used to read the result of the CALC 2 operation. If CALC 2 is disabled or
NONE is selected, the "raw" reading is read instead.
Note that this command does not initiate a calculation. This command is used only to read the result
of the last calculation.
Also see
None
CALCulate3
This Calculate subsystem is used to configure and control the limit tests (LIMIT 1 test and LIMIT 2 test).

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
Input Impedance>200 TΩ
Electrometer TypeElectrometer/High Resistance Meter
Current Measurement Range20 fA to 20 mA
DisplayVacuum Fluorescent Display
Communication InterfacesIEEE-488 (GPIB), RS-232
Power Supply100 V / 120 V / 220 V / 240 V; 50 Hz / 60 Hz
Resistance Measurement RangeUp to 200 TΩ

Related product manuals