Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-161
:MDELay <NRf>
This configuration command is used to specify the measure time for the specified test.
Usage
:TSEQuence:DLEakage:MDELay <NRf>
:TSEQuence:RVCoefficient:MDELay[1] <NRf>
:TSEQuence:RVCoefficient:MDELay2 <NRf>
:MDELay?
<NRf>
0 to 99999.9 Measure delay in seconds (resistor voltage coefficient test)
0 to 10000 Delay in seconds (diode leakage test)
Details
This command is used to specify the measure time for the specified test. Note that the resistor
voltage coefficient test uses two measure delays; delay 1 and delay 2.
This configuration command is used for the following tests:
Diode leakage test
Resistor voltage coefficient test
:TSEQuence:DLEakage:MDELay <NRf> specifies the measure delay for a diode leakage test.
:TSEQuence:RVCoefficient:MDELay[1] <NRf> specifies measure delay 1 for the resistor
voltage coefficient test.
:TSEQuence:RVCoefficient:MDELay2 <NRf> specifies measure delay 2 for the resistor voltage
coefficient test.
:MDELay? queries the measure delay.
Also see
None

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
Input Impedance>200 TΩ
Electrometer TypeElectrometer/High Resistance Meter
Current Measurement Range20 fA to 20 mA
DisplayVacuum Fluorescent Display
Communication InterfacesIEEE-488 (GPIB), RS-232
Power Supply100 V / 120 V / 220 V / 240 V; 50 Hz / 60 Hz
Resistance Measurement RangeUp to 200 TΩ