EasyManuals Logo
Home>Keithley>Measuring Instruments>6517B

Keithley 6517B Reference Manual

Keithley 6517B
498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #399 background imageLoading...
Page #399 background image
Section 14: SCPI command reference Model 6517B Electrometer Reference Manual
14-148 6517B-901-01 Rev. C / August 2015
:DELay <n>
This command is used to set the scan and measure layer delays.
Usage
:ARM[:SEQuence[1]]:LAYer2:DELay <n>
:TRIGger[:SEQuence[1]]:DELay <n>
:DELay?
:DELay? DEFault
:DELay? MINimum
:DELay? MAXimum
<n>
ï‚· 0 to 999999.999: Specify delay in seconds
ï‚· DEFault: 0 second delay
ï‚· MINimum: 0 second delay
ï‚· MAXimum: 999999.999 second delay
Details
These delay periods are used to delay operation in the specified layer. After the programmed event
occurs, the instrument waits until the delay period expires before proceeding on in the trigger model
(see the IEEE-488 trigger model (on page 11-16)). Note that the arm layer (Arm Layer 1) does not
use a delay.
Also see
None

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals