:DELay <n>
This command is used to set the scan and measure layer delays.
Usage
:ARM[:SEQuence[1]]:LAYer2:DELay <n>
:TRIGger[:SEQuence[1]]:DELay <n>
:DELay?
:DELay? DEFault
:DELay? MINimum
:DELay? MAXimum
<n>
0 to 999999.999: Specify delay in seconds
DEFault: 0 second delay
MINimum: 0 second delay
MAXimum: 999999.999 second delay
Details
These delay periods are used to delay operation in the specified layer. After the programmed event
occurs, the instrument waits until the delay period expires before proceeding on in the trigger model
(see the IEEE-488 trigger model (on page 11-16)). Note that the arm layer (Arm Layer 1) does not
use a delay.
Also see
None

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
Input Impedance>200 TΩ
Electrometer TypeElectrometer/High Resistance Meter
Current Measurement Range20 fA to 20 mA
DisplayVacuum Fluorescent Display
Communication InterfacesIEEE-488 (GPIB), RS-232
Power Supply100 V / 120 V / 220 V / 240 V; 50 Hz / 60 Hz
Resistance Measurement RangeUp to 200 TΩ

Related product manuals