Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-39
[:DATA] <n>
This command is used to specify LIMIT 1 and LIMIT 2 upper and lower limits.
Usage
:CALCulate3:LIMit[1]:UPPer[:DATA] <n>
:CALCulate3:LIMit[1]:LOWer[:DATA] <n>
:CALCulate3:LIMit2:UPPer[:DATA] <n>
:CALCulate3:LIMit2:LOWer[:DATA] <n>
:UPPer?
:UPPer? DEFault
:UPPer? MINimum
:UPPer? MAXimum
:LOWer?
:LOWer? DEFault
:LOWer? MINimum
:LOWer? MAXimum
<n>
-9.999999e35 to +9.999999e35: Specify limit value
DEFault: Set specified upper limit to 1; set specified lower limit to -1
MINimum: Set specified limit to -9.999999e35
MAXimum: Set specified limit to +9.999999e35
Details
These commands are used to set the upper and lower limits for LIMIT 1 and LIMIT 2. The actual limit
depends on which measurement function is currently selected. For example, a limit value of 1 is 1
A for the amps function and 1 C for the coulombs function. A limit value is not range sensitive. A
limit of 2 for volts is 2 V on all measurement ranges.
Also see
None

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals