Section 15: Calibration procedure Model 6517B Electrometer Reference Manual
15-22 6517B-901-01 Rev. C / August 2015
Voltage source calibration commands
This is a list of all voltage source calibration commands for the Model 6517B.
Voltage source calibration commands
Command Description
:CAL:PROT:VSETZ100
Start the 100 V range zero calibration
:CAL:PROT:VSRCZ100 <measured voltage>
Calibrate the 100 V range zero
:CAL:PROT:VSET100
Start the +100 V range calibration
:CAL:PROT:VSRC100 <measured voltage>
Calibrate the +100 V range
:CAL:PROT:VSETN100
Start the -100 V range calibration
:CAL:PROT:VSRCN100 <measured voltage>
Calibrate the -100 V range
:CAL:PROT:VSET40
Start the +40 V calibration
:CAL:PROT:VSRC40 <measured voltage>
Calibrate the +40 V output
:CAL:PROT:VSETZ1000
Start the 1000 V range zero calibration
:CAL:PROT:VSRCZ1000 <measured voltage>
Calibrate the 1000 V range zero
:CAL:PROT:VSET1000
Start the +1000 V range calibration
:CAL:PROT:VSRC1000 <measured voltage>
Calibrate the +1000 V range
:CAL:PROT:VSETN1000
Start the -1000 V range calibration
:CAL:PROT:VSRCN1000 <measured voltage>
Calibrate the -1000 V range
:CAL:PROT:VSET400
Start the +400 V calibration
:CAL:PROT:VSRC400 <measured voltage>
Calibrate the +400 V output
Humidity calibration commands
This is a list of all Model 6517B humidity calibration commands
Humidity calibration commands
Command Description
:CAL:PROT:HUMZERO
Calibrate the humidity input zero
:CAL:PROT:HUM1 <1 volt value>
Calibrate the humidity +1 V range