Section 14: SCPI command reference Model 6517B Electrometer Reference Manual
14-154 6517B-901-01 Rev. C / August 2015
:SSYNchronous:LINE <NRf>
This command is used to specify the semi-synchronous trigger link for the measure layer.
Usage
:TRIGger[:SEQuence[1]]:TCONfigure:SSYNchronous :LINE <NRf>
:LINE?
<NRf>
ï‚· 1: Line #1
ï‚· 2: Line #2
ï‚· 3: Line #3
ï‚· 4: Line #4
ï‚· 5: Line #5
ï‚· 6: Line #6
Details
This command is used to select one of the six trigger lines for the semi-synchronous trigger link.
Also see
None
RTCLock commands
The following commands are used to specify the date and time for the RTCLock control source (see :TSTamp
commands (on page 14-127) for more information).
:DATE <yr>, <mo>, <day>
This command is used to set the date for the RTCLock control source.
Usage
:ARM[:SEQuence[1]][LAYer[]]:RTCLock:DATE <yr>, <mo>, <day>
:DATE?
<yr>
ï‚· 2005 to 2104: Specify year
ï‚· <mo> = 1 to 12: Specify month
ï‚· <day> = 1 to 31: Specify day
Details
This command is used to set the date for the RTCLock control source. Setting an invalid date (for
example, February 30, 2007) for the clock event results in an error and the previous date is retained.
Also see
None

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals