Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-21
:STATus command summary
Command Description
Default
parameter
SCPI
:STATus

:MEASurement
Path to control measurement event registers:
[:EVENt]?
Read the event register (Note 2)
:ENABle <NRf>
Program the enable register (Note 3)
:ENABle?
Read the enable register
:PTRansition <NRf>
Program the positive transition register (Note 4)
:PTRansition?
Read the positive transition register
:NTRansition <NRf>
Program the negative transition register (Note 5)
:NTRansition?
Read the negative transition register
:CONDition?
Read the condition register
:OPERation

[:EVENt]?
Read the event register (Note 2)

:ENABle <NRf>
Program the enable register (Note 3)

:ENABle?
Read the enable register

:PTRansition <NRf>
Program the positive transition register (Note 4)

:PTRansition?
Read the positive transition register

:NTRansition <NRf>
Program the negative transition register (Note 5)

:NTRansition?
Read the negative transition register

:CONDition?
Read the condition register

:ARM
Path to control arm event registers:

[:EVENt]?
Read the event register (Note 2)

:ENABle <NRf>
Program the enable register (Note 3)

:ENABle?
Read the enable register

:PTRansition
Program the positive transition register (Note 4)

:PTRansition?
Read the positive transition register

:NTRansition <NRf>
Program the negative transition register (Note 5)

:NTRansition?
Read the negative transition register

:CONDition?
Read the condition register

:SEQuence
Path to control sequence status registers

[:EVENt]?
Read the event register (Note 2)

:ENABle <NRf>
Program the enable register (Note 3)

:ENABle?
Read the enable register

:PTRansition <NRf>
Program the positive transition register (Note 4)

:PTRansition?
Read the positive transition register

:NTRansition <NRf>
Program the negative transition register (Note 5)

:NTRansition?
Read the negative transition register

:CONDition?
Read the condition register

:STATus
:OPERation
:TRIGger
Path to control trigger status registers:

[:EVENt]?
Read the event register (Note 2)

:ENABle <NRf>
Program the enable register (Note 3)

:ENABle?
Read the enable register

:PTRansition <NRf>
Program the positive transition register (Note 4)


Table of Contents

Other manuals for Keithley 6517B