Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-163
:STIMe <NRf>
This command specifies the bias time for the specified test.
Usage
:TSEQuence:SRESistivity:STIMe <NRf>
:TSEQuence:VRESistivity:STIMe <NRf>t
:TSEQuence:SIResistance:STIMe <NRf>
:TSEQuence:STSWeep:STIMe <NRf>
:STIMe?
<NRf>
0 to 99999.9 delay in seconds (staircase sweep test)
0 to 99999.9 Delay in seconds (all other tests)
Details
This command is used to specify the bias time for the specified test.
This configuration command is used for the following tests:
ï‚· Surface resistivity test
ï‚· Volume resistivity test
ï‚· Surface insulation resistance test
ï‚· Staircase sweep test
:TSEQuence:SRESistivity:STIMe specifies the bias time for the surface resistivity test.
:TSEQuence:VRESistivity:STIMe specifies the bias time for the volume resistivity test.
:TSEQuence:SIResistance:STIMe specifies the bias time for the surface insulation resistance
test.
:TSEQuence:STSWeep:STIMe specifies the bias time for the staircase sweep test
:STIMe? queries the bias time.
Also see
None

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals