Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-27
Command Description Default parameter SCPI
:DELay <n>
Program delay (0 to 999999.999 sec) 0

:DELay?
Query delay

:SOURce <name>
Select control source (HOLD,
IMMediate, TIMer, MANual, BUS,
TLINk, or EXTernal)
IMMediate
(Note 2)
:SOURce?
Query control source
:TIMer <n>
Set timer interval (0 to 999999.999
sec)
0.1

:TIMer?
Query timer

:SIGNal
Loop around control source

:TCONfigure
Path to configure triggers:

:DIRection
<name>
Enable (SOURce) or disable
(ACCeptor) Bypass
ACCeptor
:DIRection?
Query direction
:ASYNchronous
Path to configure asynchronous
trigger link:
:ILINe
<NRf>
Select input line (1 to 6) 2
:ILINe?
Query input line
:OLINe
<NRf>
Select output line (1 to 6) 1
:OLINe?
Query output line
:TRIGger[:SEQuence[1
]]
Path to program trigger layer:

:IMMediate
Loop around control source

:COUNt <n>
Set measure count (1 to 99999, or INF) (Note 3)

:COUNt?
Query measure count

:DELay <n>
Set delay (0 to 999999.999 sec) 0

:DELay?
Query delay
:SOURce <name>
Select control source (HOLD,
IMMediate, TIMer, MANual, BUS,
TLINk, or EXTernal)
IMMediate
(Note 2)
:SOURce?
Query control source

:TIMer <n>
Set timer interval (0 to 999999.999 sec) 0.1

:TIMer?
Request the programmed timer interval

:SIGNal
Loop around control source

:TCONfigure
Path to configure triggers:
:PROTocol
<name>
Select protocol (ASYNchronous or
SSYNchronous)
ASYN
:PROTocol?
Query protocol
:DIRection
<name>
Enable (SOURce) or disable
(ACCeptor) Bypass
ACCeptor
:DIRection?
Query direction
:ASYNchronous
Path to configure asynchronous
trigger link: