EasyManua.ls Logo

Keithley 6517B

Keithley 6517B
498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 14: SCPI command reference Model 6517B Electrometer Reference Manual
14-28 6517B-901-01 Rev. C / August 2015
Command Description Default parameter SCPI
:ILINe <NRf>
Select input line (1 to 6) 2
:ILINe?
Query input line
:OLINe <NRf>
Select output line (1 to 6) 1
:OLINe?
Query output line
:SSYNchronous
Path to configure semi-synchronous
trigger link:
:LINE <NRf>
Select trigger line (1 to 6) 1
:LINE?
Query trigger line
Notes:
1. Defaults for continuous initiation:
:SYSTem:PRESet enables continuous initiation.
*RST disables continuous initiation.
2. TLINK and RTCLock (Arm Layer 1) are non-SCPI parameters for the :SOURce command. The
command (:SOURce), and all the other parameters are SCPI confirmed.
3. Defaults for count (Arm Layer 2 and trigger layer):
:SYSTem:PRESet sets the count to INF (infinite).
*RST sets the count to 1.

Table of Contents

Other manuals for Keithley 6517B

Related product manuals