Section 7: Triggering Model 6517B Electrometer Reference Manual
7-20 6517B-901-01 Rev. C / August 2015
The trigger link connections for this test system are shown in the following figure. The trigger link of
the Model 6517B is connected to trigger link of the Model 7001/7002 Switch System. Notice that only
one trigger link cable is needed. For this example, the Models 6517B and 7001/7002 are configured
as follows:
Figure 83: Trigger link connections (asynchronous example)
Model 7001 or 7002:
Idle State:
Reset = :INIT:CONT OFF*
Scan List = 1!1-1!10,
Arm layer:
Arm spacing = Immediate*
Arm count = 1*
Arm trigger control = Acceptor*
Scan layer:
Scan spacing = Immediate*
Number of scans = 1
Scan trigger control = Acceptor*
Channel Layer:
Channel spacing = TrigLink
Trigger link mode = Asynchronous
Input line = #4
Output line = #3
Number of channels = 20
Channel trigger control = source*
* Indicates that the setting is the RESET (and factory) default condition.

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
Input Impedance>200 TΩ
Electrometer TypeElectrometer/High Resistance Meter
Current Measurement Range20 fA to 20 mA
DisplayVacuum Fluorescent Display
Communication InterfacesIEEE-488 (GPIB), RS-232
Power Supply100 V / 120 V / 220 V / 240 V; 50 Hz / 60 Hz
Resistance Measurement RangeUp to 200 TΩ

Related product manuals