Section 10: Limits, digital I/O, and scanning Model 6517B Electrometer Reference Manual
10-2 6517B-901-01 Rev. C / August 2015
The limit test is performed after MATH operations. Unit prefixes are applied before the limit test, for
example:
Low Limit = -10.0, High limit = 10.0. A 1.5V reading passes (PASS)
Low Limit = -10.0, High limit = 10.0. A 0.6 M reading equals 600,000 (FAIL)
Limit set #1 is available as a multiple display, which shows a bar graph and a pass/fail indication for
the measurements. While the limits bar graph is displayed, the programmed values for limit set #1
can be viewed by pressing INFO.
The digital output lines can be programmed to reflect the results of limit tests. The first test that fails
sets a programmable digital output pattern, where the test execution order is LOW1, HIGH1, LOW2,
HIGH2. If all tests pass, another programmable pattern is set.

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
Input Impedance>200 TΩ
Electrometer TypeElectrometer/High Resistance Meter
Current Measurement Range20 fA to 20 mA
DisplayVacuum Fluorescent Display
Communication InterfacesIEEE-488 (GPIB), RS-232
Power Supply100 V / 120 V / 220 V / 240 V; 50 Hz / 60 Hz
Resistance Measurement RangeUp to 200 TΩ

Related product manuals