EasyManuals Logo
Home>Keithley>Measuring Instruments>6517B

Keithley 6517B Reference Manual

Keithley 6517B
498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #265 background imageLoading...
Page #265 background image
Section 14: SCPI command reference Model 6517B Electrometer Reference Manual
14-14 6517B-901-01 Rev. C / August 2015
Command Description Default parameter SCPI
:AUTO?
Query auto resolution
:DIGits?
Query resolution
:AVERage
Path to control the average filter:
:TYPE <name>
Select filter type (NONE, SCALar,
ADVanced)
SCALar
:TYPE?
Query filter type
:TCONtrol <name>
Select filter type: (MOVing or REPeat)
(Note 3)
:TCONtrol?
Query filter type
:COUNt <n>
Specify filter count (1 to 100) 10
:COUNt?
Query filter count
:ADVanced
Path to configure and control advanced
filter
:NTOLerance <n>
Specify noise tolerance level (0 to 100 %) 1
:NTOLerance?
Query noise tolerance level
[:STATe] <b>
Enable or disable digital filter OFF
[:STATe]?
Query state of digital filter
:MEDian
Path to control the median filter: ON
[:STATe] <b>
Enable or disable median filter
[:STATe]?
Query state of median filter
:RANK <NRf>
Specify rank for median filter (1 to 5) 1
:RANK?
Query rank
:GUARd <b>
Enable or disable guard OFF
:GUARd?
Query state of guard
:XFEedback <b>
Enable or disable external feedback OFF
:XFEedback?
Query state of external feedback
:CURRent[:DC]
Path to configure Amps:

:APERture <n>
Set integration rate in seconds (166.67e-6
to 200e-3)
(Note 1)
:AUTO <b>
Enable or disable auto aperture OFF

:AUTO ONCE
Enable and then disable auto aperture
:AUTO?
Query auto aperture
:APERture?
Query aperture (integration rate)

:NPLCycles <n>
Set integration rate (line cycles; 0.01 to
10)
1

:AUTO <b>
Enable or disable auto NPLC OFF
:AUTO ONCE
Enable and then disable auto NPLC
:AUTO?
Query auto line cycle integration
:NPLCycles?
Query line cycle integration rate

:RANGe
Path to configure measurement range:

[:UPPer] <n>
Select range (0 to 21e-3) 20e-3

[:UPPer]?
Query range

:AUTO <b>
Enable or disable auto range (Note 4)

:AUTO ONCE
Set range based on present input signal


Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals