348 Keysight N9927-90001 User’s Guide
5G Rapid Prototype Phased Array Antenna – Option 360
How to Set Up Phased Array Antenna Settings
11. Select Trace 6—choose:
For this step, refer to Figure 6 on page 337.
Polar Size Max Med Min
Sets the size of the polar display. Smaller
sizes used to overlay PNG maps. (Default = Max)
Polar Orientation
Aligns the antenna polar display to a selected
direction.
Image Enable OFF ON
(Optional to overlay antenna bore-sight
direction. Only active if you have recalled a PNG image to display on
the FieldFox as a map. (Default = OFF)
Recall Image
Use to recall PNG map to overlay antenna direction.
Compass Image OFF ON
To display a compass indicating true north
for antenna. (Default = ON)
Compass North
Use this softkey to align the compass’ true north
value in degrees.
Repeat process as needed for additional settings.
12. Optional: Press Freq/Dist
Press Antenna Settings
opens the table antenna settings softkey
menu, where you can define your phased array antenna settings,
Ball Antenna: LAN address, Settings: Resolution BW, Video BW,
Beam Width, Array Antenna: Center frequency, Azimuth Ref,
Elevation Ref, and Polar Display: Ref. For details, refer to step 6
above.
Then Start Azimuth
+/-45 degrees. (Default = 45 degrees)
Then Stop Azimuth
+/-45 degrees. (Default = 45 degrees)
Then Ref Elevation
+/-45 degrees. (Default = 0 degrees)
13. Optional: choose change sweep, dwell time, and max hold settings:
Press Sweep
Then choose:
Dwell Time
to set the delay between the antenna positioning from
1 ms to 10 seconds (default: 10 ms). For pulsed RF measurements it
is recommended that you increase the Dwell Time to ~100 ms.
Max Hold
For pulsed RF measurements, it is recommended that
Max Hold is set to OFF when using the Heat Map display.
(Default = OFF)
Single
Continuous
to enable or disable the continuous sweep (default:
ON).

Table of Contents

Other manuals for Keysight Technologies FieldFox N9938A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies FieldFox N9938A and is the answer not in the manual?

Keysight Technologies FieldFox N9938A Specifications

General IconGeneral
BrandKeysight Technologies
ModelFieldFox N9938A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals