Example 4 Create a trigger source that triggers a measurement 50 milliseconds after the output has
settled..
Front Panel Menu Reference SCPI Command
Select System\Signal\Define.
Select Expression 4 in the dropdown list.
Enter "Delay(OutpSettled, 0.05)" in the text field.
You can enter any value from the numeric keypad. For additional characters, use
the up/down navigation keys to enter an alpha character by scrolling through the
selection list that appears when you press the keys. Use the left/right navigation
keys to traverse the text field. Use the backspace key to delete a value. Press Enter
when you are finished.
Program the signal expression.
SYST:SIGN:DEF EXPR4, "Delay
(OutpSettled,0.05)"
Program the measurement trigger source.
TRIG:ACQ:SOUR EXPR4
4 Using the Advanced Power System
150 Keysight N6900/N7900 Series Operating and Service Guide