EasyManuals Logo

Keysight Technologies N6900 Series User Manual

Keysight Technologies N6900 Series
358 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #217 background imageLoading...
Page #217 background image
CALibrate:SAVE
Saves the calibration constants in non-volatile memory. Do this at the end of the calibration to avoid
losing changes.
Parameter Typical Return
(none) (none)
Store calibration constants into non-volatile memory: CAL:SAVE
CALibrate:STATe 0|OFF|1|ON [,<password>]
CALibrate:STATe?
Enables or disables calibration mode. Calibration mode must be enabled for the instrument to accept
any calibration commands. The first parameter specifies the state. The second optional parameter is
the password.
Parameter Typical Return
0|OFF|1|ON, *RST OFF 0or1
<password>a numeric value up to 15 digits (none)
Disable calibration: CAL:STAT OFF
Enable calibration: CAL:STAT ON [,value]
<password> is required if it has been set to a non-zero value.
CALibrate:VOLTage[:LEVel] <value>
Calibrates the local voltage programming and measurement. The value selects the range to
calibrate.
Parameter Typical Return
The maximum voltage of the output range. (none)
Calibrates the voltage of the 20 V range: CAL:VOLT20
CALibrate:VOLTage:CMRR
Calibrates the voltage common mode rejection ratio.
Parameter Typical Return
(none) (none)
Calibrates the common mode rejection ratio: CAL:VOLT:CMRR
6 SCPI Programming Reference
217 Keysight N6900/N7900 Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N6900 Series and is the answer not in the manual?

Keysight Technologies N6900 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelN6900 Series
CategoryPower Supply
LanguageEnglish

Related product manuals