Parameter Typical Return
(none) <points>
Returns the number of points in the dwell list: LIST:DWEL:POIN?
[SOURce:]LIST:STEP ONCE|AUTO
[SOURce:]LIST:STEP?
Specifies how the list responds to triggers. ONCE causes the output to remain at the present step until
a trigger advances it to the next step. Triggers that arrive during the dwell time are ignored. AUTO
causes the output to automatically advance to each step, after the receipt of an initial starting trigger.
Steps are paced by the dwell list. As each dwell time elapses, the next step is immediately output.
Parameter Typical Return
ONCE|AUTO, *RST AUTO ONCE or AUTO
Specifies the list steps to be paced by trigger signals: LIST:STEP ONCE
[SOURce:]LIST:TERMinate:LAST 0|OFF|1|ON
[SOURce:]LIST:TERMinate:LAST?
Determines the output value when the list terminates. When ON (1), the output voltage or current
remains at the last list step. The value of the last voltage or current list step becomes the IMMediate
value when the list completes. When OFF (0), and also when the list is aborted, the output returns to
the settings that were in effect before the list started.
Parameter Typical Return
0|OFF|1|ON , *RST OFF 0or1
Terminatewith the output atthe last step value: LIST:TERM:LAST ON
[SOURce:]LIST:TOUTput:BOSTep[:DATA] 0|OFF|1|ON {,0|OFF|1|ON }
[SOURce:]LIST:TOUTput:BOSTep[:DATA]?
[SOURce:]LIST:TOUTput:EOSTep[:DATA] 0|OFF|1|ON {,0|OFF|1|ON }
[SOURce:]LIST:TOUTput:EOSTep[:DATA]?
Specifies which list steps generate a trigger signal at the beginning of step (BOSTep) or end of step
(EOSTep). A trigger is only generated when the state is set to ON. The trigger signal can be used as a
trigger source for measurements and transients of other units, and for digital port pins configured as
trigger outputs.
Parameter Typical Return
0|OFF|1|ON 0or1
To generatetriggers at the beginning of the second step of a 3-step list:
LIST:TOUT:BOST OFF,ON,OFF
Keysight N6900/N7900 Series Operating and Service Guide 242
6 SCPI Programming Reference