:POWer
[:LEVel] <value> Sets the power level for comparator 1,2,3, or 4.
:VOLTage
[:LEVel] <value> Sets the voltage level for comparator 1,2,3, or 4.
:WHOur
[:LEVel] <value> Sets the watt-hour level for comparator 1,2,3, or 4.
:WHOur
:RESet Resets the accumulated watt-hour measurement.
:WINDow
[:TYPE] HANNing|RECTangular Selects the measurement window.
SENSe:AHOur:RESet
SENSe:WHOur:RESet
Resets the amp-hour or watt-hour measurement to zero.
Parameter Typical Return
(none) (none)
Resets the amp-hour measurement: SENS:AHO:RES
SENSe:BBR:PERiod <value>
SENSe:BBR:PERiod?
Sets the recording period of the black box recorder in seconds. This parameter is saved in non-volatile
memory.
Parameter Typical Return
0.01 | 0.1 <period>
Selects 0.01 seconds: SENS:BBR:PER 0.01
SENSe:CURRent[:DC]:RANGe[:UPPer] <value>|MIN|MAX
SENSe:CURRent[:DC]:RANGe[:UPPer]? [MIN|MAX]
Selects a DC current measurement range. The value that you enter must be the highest value that
you expect to measure. Units are in amperes. The instrument selects the range with the best
resolution for the value entered.
Parameter Typical Return
0 to maximum <max current>
Selects the 2 A range: SENS:CURR:RANG 2
l The MAX query returns the maximum DC current that can be measured on the present range.
6 SCPI Programming Reference
257 Keysight N6900/N7900 Series Operating and Service Guide