SENSe:CURRent[:DC]:RANGe:AUTO 0|OFF|1|ON
SENSe:CURRent[:DC]:RANGe:AUTO?
Enables or disables seamless measurement auto ranging.
Parameter Typical Return
0|OFF|1|ON, *RST ON 0or1
Enable current measurement autoranging: SENS:CURR:RANG:AUTO ON
SENSe:ELOG:CURRent[:DC]:RANGe[:UPPer] <value>|MIN|MAX
SENSe:ELOG:CURRent[:DC]:RANGe[:UPPer]? [MIN|MAX]
Selects the Elog current measurement range. The value that you enter must be the highest value that
you expect to measure. Units are in amperes. The instrument selects the range with the best
resolution for the value entered.
Parameter Typical Return
0 to maximum <max current>
Selects the 1 A range: SENS:ELOG:CURR:RANG 1
l The MAX query returns the maximum DC current that can be measured on the present range.
SENSe:ELOG:CURRent[:DC]:RANGe:AUTO 0|OFF|1|ON
SENSe:ELOG:CURRent[:DC]:RANGe:AUTO?
Enables or disables Elog seamless measurement autoranging.
Parameter Typical Return
0|OFF|1|ON, *RST ON 0or1
Enable Elog current autoranging: SENS:ELOG:CURR:RANG:AUTO ON
SENSe:ELOG:FUNCtion:CURRent 0|OFF|1|ON
SENSe:ELOG:FUNCtion:CURRent?
SENSe:ELOG:FUNCtion:VOLTage 0|OFF|1|ON
SENSe:ELOG:FUNCtion:VOLTage?
Enables or disables the Elog current or voltage measurement function.
Parameter Typical Return
0|OFF|1|ON, *RST OFF 0or1
Enables datalog current measurements: SENS:ELOG:FUNC:CURR ON
Keysight N6900/N7900 Series Operating and Service Guide 258
6 SCPI Programming Reference