Ch.3 Sequence of Examination
< 51 >
REGIUS Console CS-3 OPERATION MANUAL Ver. 1.10 2005.2
When a post registered cassette is being used and an image is read from a cassette differing from that in the order,
the following procedure may be followed to substitute the image into the correct order.
Image Substitution
1 Following the image erasing procedure, select
the order mistakenly read and touch the [NG]
button.
•A message requesting confirmation of erasure of
data read will be displayed.
2 Touch the [Recovery] button.
• Orders for which the read process has not been
executed will be displayed on an order selection
list.
Example
The image read into the "skull A-
>P" order has been mistakenly read
into the "skull R->L" order.
The image substitution
can be carried out only
when the image is read
by the cassette reader
with a screen menu
selection method.
p.49 Image
Erasing
3 Select the correct order for the image read.
• The selected order will be displayed in a green
frame.
4 Touch the [OK] button.
• Display will return to the examination search
screen.
• The image of the order selected in step
1 will be
displayed in the image display field of the order
selected in step
3.
Selection of the "skull A->P" order.
Only cassette reader
orders images of which
have not been read are
displayed as orders
available for substitu-
tion on the order selec-
tion screen.
Transfer of the image to the "skull
A->P" order.
1
2
3
4