EasyManuals Logo
Home>Mindray>Medical Equipment>A5

Mindray A5 Operating Instructions

Mindray A5
310 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #111 background imageLoading...
Page #111 background image
A5/A3™ Operating Instructions 046-003777-00 4 - 7
Preoperative Tests System Self-Test
RESULT COMMENTS/OPTIONS
Pass condition
Example:
The Pass condition indicates that the A5/A3 has passed
the System Self-Test. No errors have been detected. Alarms
and hardware are functioning properly.
Select Continue to enter the Automatic Circuit Leak and
Compliance Test screen.
or
Select Skip to enter the Standby with automatic
ventilation enabled.
All-Functional error condition
Example:
The All-Functional error condition indicates that errors
have been detected. However, all automatic ventilation,
manual, and bypass modes are still enabled.
Select Accept to enter the Automatic Circuit Leak and
Compliance Test screen.
or
Select Skip to enter the main screen with automatic
ventilation enabled,
Manual Only error condition
Example:
The Manual Only error condition indicates that the A5/A3
can be used in manual mode only.
Select Retry to repeat the System Self-Test.
or
Select Manual Only to place the device in manual
ventilation mode only. The following low priority alarm
will be displayed on the main screen: Automatic
Ventilation Disabled.
WARNING: Selecting the “Manual
Only” button will
disable automatic
ventilation.
TABLE 4-2 Types of System Self-Test Results

Table of Contents

Other manuals for Mindray A5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A5 and is the answer not in the manual?

Mindray A5 Specifications

General IconGeneral
BrandMindray
ModelA5
CategoryMedical Equipment
LanguageEnglish

Related product manuals