System Self-Test Preoperative Tests
4 - 8 046-003777-00 A5/A3â„¢ Operating Instructions
Bundle Version – The Bundle Version is displayed in all System Self-Test results. The Bundle Version is
the version number of the package of software that is installed in the A5/A3. If the Bundle Version
displays a fail status, contact Mindray Technical Support.
Machine Non-Functional error condition
Example:
The Machine Non-Functional error condition indicates
that the A5/A3 cannot be used.
Select Retry to repeat the System Self-Test.
or
Contact service if this error condition persists.
NOTE: The Service Access
button is only available
to Mindray-authorized
service personnel and
requires a service
password.
RESULT COMMENTS/OPTIONS
TABLE 4-2 Types of System Self-Test Results