A5/A3™ Operating Instructions 046-003777-00 C - 11
Parameters and Factory Defaults Ventilation Modes
Tinsp — Range: 0.2 to 5 sec
Step: 0.1 sec
Defaults:
Adult: 2.0 sec
Pediatric: 1.0 sec
Infant: 1.0 sec
— Range: 0.2 to 5 sec
Step: 0.1 sec
Defaults:
Adult: 2.0 sec
Pediatric: 1.0 sec
Infant: 1.0 sec
— —
Pinsp — — Range:
PEEP+5 to 70 cmH
2
O
Step: 1 cmH
2
O
Defaults:
Adult: 15 cmH
2
O
Pediatric: 10 cmH
2
O
Infant: 10 cmH
2
O
Range:
PEEP+5 to 70 cmH
2
O
Step: 1 cmH
2
O
Defaults:
Adult: 15 cmH
2
O
Pediatric: 10 cmH
2
O
Infant: 10 cmH
2
O
— —
Tpause Range:
Off, 5% to 60%
Step: 1%
Default: 10%
Range:
Off, 5% to 60%
Step: 1%
Default: 10%
— — — —
Plimit Range:
10 to 100 cmH
2
O
Step: 1 cmH
2
O
Defaults:
Adult: 50 cmH
2
O
Pediatric: 40 cmH
2
O
Infant: 20 cmH
2
O
Range:
10 to 100 cmH
2
O
Step: 1 cmH
2
O
Defaults:
Adult: 50 cmH
2
O
Pediatric: 40 cmH
2
O
Infant: 20 cmH
2
O
— — — —
PlimVG (A5
only)
— — Range: 5 - 100
cmH
2
O
Step: 1 cmH
2
O
Default:
Pinsp
— — —
PEEP Range:
Off, 3 to 30 cmH
2
O
Step: 1 cmH
2
O
Default: Off
Range:
Off, 3 to 30 cmH
2
O
Step: 1 cmH
2
O
Default: Off
Range:
Off, 3 to 30 cmH
2
O
Step: 1 cmH
2
O
Default: Off
Range:
Off, 3 to 30 cmH
2
O
Step: 1 cmH
2
O
Default: Off
Range:
Off, 3 to 30 cmH
2
O
Step: 1 cmH
2
O
Default: Off
—
PARAMETER VCV SIMV-VC PCV SIMV-PC PS MANUAL