Ventilation Modes Parameters and Factory Defaults
C - 12 046-003777-00 A5/A3™ Operating Instructions
ΔP — Range:
3 to 50 cmH
2
O
Step: 1
Defaults:
Adult: 8 cmH
2
O
Pediatric: 5 cmH
2
O
Infant: 5 cmH
2
O
— Range:
3 to 50 cmH
2
O
Step: 1
Defaults:
Adult: 8 cmH
2
O
Pediatric: 5 cmH
2
O
Infant: 5 cmH
2
O
Range:
3 to 50 cmH
2
O
Step: 1
Defaults:
Adult: 8 cmH
2
O
Pediatric: 5 cmH
2
O
Infant: 5 cmH
2
O
—
Trigger — Range:
1 to 15 L/min
Step: 1
Defaults:
Adult: 3 L/min
Pediatric: 2 L/min
Infant: 2 L/min
— Range:
1 to 15 L/min
Step: 1
Defaults:
Adult: 3 L/min
Pediatric: 2 L/min
Infant: 2 L/min
Range:
1 to 15 L/min
Step: 1
Defaults:
Adult: 3 L/min
Pediatric: 2 L/min
Infant: 2 L/min
—
Tslope * — Range: 0.0 to 2.0 sec
Step: 0.1 sec
Default: 0.2 sec
Range: 0.0 to 2.0 sec
Step: 0.1 sec
Default: 0.2 sec
Range: 0.0 to 2.0 sec
Step: 0.1 sec
Default: 0.2 sec
Range: 0.0 to 2.0 sec
Step: 0.1 sec
Default: 0.2 sec
—
PS — Range: On, Off
Step: —
Default: Off
— Range: On, Off
Step: —
Default: Off
— —
Bypass (A5
only)
— — — — — Range: On, Off
Step: —
Default: Off
Alarm — — — — — Range: On, Off
Step: —
Default: On
Apnea Ti — — — — Range: 0.2 to 5.0 sec
Step: 0.1 sec
Default:
5.0 sec (adult)
3.0 sec (Pediatric)
2.0 sec (Infant)
—
* The Tslope setting is an approximation. The exact waveform shape may not be realized under certain clinical scenarios.
PARAMETER VCV SIMV-VC PCV SIMV-PC PS MANUAL