A5/A3™ Operating Instructions 046-003777-00 3 - 39
System Interface System Tab
SYSTEM TAB
BUTTON CHOICES DESCRIPTION
Calibration External AG Module
O
2
Sensor
Select to calibrate the External AG
Module
or O
2
sensor. Follow the screen
instructions. The date and time of the
last calibration is displayed next to the
O
2
Sensor or External AG Module
button.
NOTE: The AG module information
appears only when an AG module is
connected to the A5 system.
Language ENGLISH (default)
CHINESE
FRENCH
SPANISH
Select to set the user interface text
language.
Default Settings Default Patient Size
(default=Infant, Adult, Pediatric)
Default Vent Mode (default=VCV,
SIMV-VC, PCV, SIMV-PC, PS)
NOTE: Default changes take
effect after next case
or when O.R.
defaults are loaded.
Select to set the default patient size.
Select to set the default mechanical
ventilation mode.
For default changes to take effect:
1. Press Accept.
2. Start next case.
3. End case.
Manage Defaults Save Defaults
Save as O.R. Defaults
Load User Defaults
Load O.R. Defaults
Restore Partial Defaults
Import Defaults
Export Defaults
Select “Save Defaults” or “Save as O.R.
Defaults” to save the current
configuration as the user default
configuration.
Select “Load User Defaults” or “Load
O.R. Defaults” to load the user default
configuration.
Select “Restore Partial Defaults” to
overwrite the user defaults and system
settings with the factory default
settings. Note that network settings
will not be restored.
Select “Import Defaults” to import a
copy of the defaults from the USB mass
storage device if one has been inserted
into an SB port at the rear of the A5/A3
unit.
Select “Export Defaults” to export a
copy of the defaults to the USB mass
storage device if one has been inserted
into an SB port at the rear of the A5/A3
unit.
TABLE 3-10 System Tab Settings