A-14
*Measurement accuracy verification: In adult and pediatric modes, the blood pressure measurements measured with this
device are in compliance with the Standard for Non-invasive sphygmomanometers (ISO 81060-2) in terms of mean error and
standard deviation by comparing with intra-arterial or auscultatory measurements (depending on the configuration) in a
typical patient population. For auscultatory reference, the 5th Korotkoff sound was used to determine the diastolic pressure.
In neonatal mode, the blood pressure measurements measured with this device are in compliance with the Standard for
Non-invasive sphygmomanometers (ISO 81060-2) in terms of mean error and standard deviation by comparing with
intra-arterial measurements (depending on the configuration) in a typical patient population.
A.8.6 Temp
Standards Meet standard of ISO 80601-2-56
Technique Thermal resistance
Operating mode Direct mode
Measurement range 0 to 50 ℃ (32 to 122 ℉)
Resolution 0.1 ℃
Accuracy ±0.1 ℃ (without probe)
Refreshing rate 1 s
Response time <5 s
Minimum time for accurate
measurement
Body surface: <100 s
Body cavity: <80 s
Minimum time between
measurements
Body surface probe: <100 s
Body cavity probe: <80 s
Alarm limit Range Step
T1/T2 High (low limit +1) to 50 ℃
0.1 ℃
T1/T2 Low 0.1 to (high limit -1) ℃
TD High 0.1 to 50 ℃