A-15
A.8.7 IBP
Standards Meet standard of IEC60601-2-34.
Technique Direct invasive measurement
IBP
Measurement range -50 to 360 mmHg
Resolution 1 mmHg
Accuracy ±2% or ±1 mmHg, whichever is greater (without sensor)
Refreshing rate 1 s
Pressure transducer
Excitement voltage 5 VDC, ±2%
Sensitivity 5 V/V/mmHg
Zero adjustment range
±200 mmHg
Impedance range 300 to 3000Ω
Alarm limit Range (mmHg) Step (mmHg)
Sys High
(low limit + 2) to 360
1
Mean High
Dia High
Sys Low
-50 to (high limit – 2) Mean Low
Dia Low