EasyManua.ls Logo

Mindray BeneView T1 - Page 264

Mindray BeneView T1
286 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C-12
Outdoor Mode
Item Name
Configurable
Default
In Config Mode In Monitor Mode
Measurement Color No Yes White
Brightness No Yes 10
Key Volume No Yes 5
Alm Volume No Yes 5
Reminder Tone No Yes High
QRS Volume No Yes 5
C.2.4 Review
Item Name
Configurable
Default
In Config Mode In Monitor Mode
Tabular Trends
Interval No Yes
General: 30 min
OR: 5 min
ICU, NICU, CCU: 30 min
Trend Group No Yes Standard
Graphic Trends
Trend Group No Yes Standard
Zoom No Yes 90 min
Waves No Yes 2
Full Disclosure
Save Waves No Yes Save ECG1 by default.
Gain No Yes x 1
Sweep No Yes 25 mm/s
C.2.5 Event
Item Name Configurable Default
In Config Mode In Monitor Mode
Waveform 1 No Yes II
Waveform 2 No Yes General, OR, ICU: I
NICU: Pleth
CCU: I
Waveform 3 No Yes General, OR, ICU: Pleth
NICU: Resp
CCU: Pleth

Table of Contents