D-6
Measurement Alarm message L I Cause and solution
CO
2
FilterLine Occluded L
C
The airway or watertrap was occluded. Check the
airway and remove the occlusion.
CO
2
No Watertrap L B Check the watertrap connections.
CO
2
Check Adapter L A There is a problem with the airway adapter. Check,
clean or replace the adapter.
CO
2
FilterLine Err L C Check if there is a leak in the CO
2
sample line or the CO
2
sample line has been occluded.
CO
2
Zero Failed L A Check the CO
2
connections. After the sensors
temperature becomes stabilized, perform a zero
calibration again.
CO
2
System Err L A Re-plug the module or restart the monitor.
CO
2
Check Cal. L C Perform a calibration.
CO
2
Check Airway L C An error occurred to the airway.
CO
2
No Filterline L A Make sure that the filterline is connected.
CO
2
No Sensor L A Make sure that the sensor is connected.
CO
2
Main Board Err H C There is a problem with the CO
2
module. Re-plug the
module or restart the monitor.
CO
2
Checking Sensor L C
CO
2
Replace
Scrubber&Pump
L C
CO
2
15V Overrange H C
CO
2
Hardware Err H C
Power 12V Too High H C There is a problem with the system power supply.
Restart the monitor.
12V Too Low H C
5V Too High H C
5V Too Low H C
3.3V Too High H C
3.3V Too Low H C
No Battery H C 1. Battery is not installed or poor contact: properly
install the battery.
2. Battery circuit failure or battery failed: contact your
service personnel.
Low Battery M C Connect the monitor to external power source, or
charge the battery using a battery charger.
Battery Depleted H C Connect the monitor to external power source, or
charge the battery using a battery charger.
T1 battery to be protected
and not work.
H C T1 battery will be soon protected and will not supply
power. If you are going to use T1 for patient transport,
please replace the battery.
T1 battery aged. Replace the
battery
L C T1 battery lifetime is expired. Replace the battery with
a new one.
Power Board Comm Err H C Restart the monitor. If the problem persists, contact
your service personnel.
RT Clock Need Reset L C Internal backup battery cell fails. Contact your service
personnel.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneView T1 and is the answer not in the manual?

Mindray BeneView T1 Specifications

General IconGeneral
BrandMindray
ModelBeneView T1
CategoryMedical Equipment
LanguageEnglish

Related product manuals