D-7
Measurement Alarm message L I Cause and solution
RT Clock Not Exist H C Contact your service personnel.
System IP Address Conflict L A Set a new IP address.
Restoring Last Config. Failed L A Restart the monitor. If the problem persists, E2PROM
may fail. Contact your service personnel.
Loading Default Config.
Failed.
L A Restart the monitor. If the problem persists, E2PROM
may fail. Contact your service personnel.
USB Drive Err M A
1. Disconnect the USB memory and reconnect it
properly.
2. If the problem persists, format the USB memory.
3. If the problem still persists, replace the USB drive.
Storage Card Err M C Restart the monitor. If the problem persists, format the
storage card.
Storage Card Space Low L A Delete unnecessary data from the storage card.
USB Drive Space Low L A Delete unnecessary data from the USB memory, or
replace the USB memory.
Read Dock E2PROM Error H C Disconnect T1 from the docking station and reconnect
it, or replace the external display.
Unknown Device L A The external device fails, or is not supported by T1, or is
not connected properly. Reconnect the external device
properly. If the problem persists, replace the external
device.
Read dock E2PROM error! H C T1does not properly connect with the T1 dock, or
unspecified external display is connected. Reconnect
T1 with T1 the docking station or replace the external
display.
Other Bed Disconnected L A Check network connection.
No CMS L A The monitor is disconnected from the CMS. Check
network connection.
PWR interrupted.
Check meas. State.
L A Power supply failed accidently. Check the
measurements when the monitor restarts.