7-8
Module Parameter
Low alarm limit High alarm limit
Auto alarm limits range
Adult/
pediatric
Neonate
Adult/
pediatric
Neonate
NIBP
NIBP-S
(SYS × 0.68 +
10) mmHg
(SYS – 15) or
45mmHg
(whichever is
greater)
(SYS × 0.86 +
38) mmHg
(SYS + 15) or
105mmHg
(whichever is
smaller)
Adult: 45 to 270
Pediatric: 45 to 185
Neonate: 40 to 115
NIBP-D
(Dia × 0.68 +
6) mmHg
(Dia – 15) or
20mmHg
(whichever is
greater)
(Dia × 0.86 + 32)
mmHg
(Dia + 15) or
80mmHg
(whichever is
smaller)
Adult: 25 to 210
Pediatric: 25 to 150
Neonate: 20 to 90
NIBP-M
(Mean × 0.68
+ 8) mmHg
(Mean – 15) or
35mmHg
(whichever is
greater)
(Mean × 0.86 +
35) mmHg
(Mean + 15 or
95)mmHg
(whichever is
smaller)
Adult: 30 to 230
Pediatric: 30 to 165
Neonate: 25 to 105
Temp
T1
(T1 – 0.5)℃ (T1 – 0.5) ℃ (T1 + 0.5)℃ (T1 + 0.5)℃ 1 to 49 ℃
T2
(T2 – 0.5)℃ (T2 – 0.5) ℃ (T2 + 0.5)℃ (T2 + 0.5)℃ 1 to 49 ℃
TD
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm limit
Same as the measurement range
IBP: ART/
Ao/
UAP/
BAP/
FAP/
LV/
P1-P4
(Arterial
pressure)
IBP-S
(SYS × 0.68+
10) mmHg
(SYS – 15) or
45mmHg
(whichever is
greater)
(SYS × 0.86+
38) mmHg
(SYS + 15) or
105mmHg
(whichever is
smaller)
Adult: 45 to 270
Pediatric: 45 to 185
Neonate: 35 to 115
IBP-D
(Dia × 0.68+
6)mmHg
(Dia – 15) or
20mmHg
(whichever is
greater)
(Dia × 0.86+
32)mmHg
(Diav15) or
80mmHg
(whichever is
smaller)
Adult: 25 to 225
Pediatric: 25 to 150
Neonate: 20 to 90
IBP-M
(Mean × 0.68
+ 8)mmHg
(Mean – 15) or
35mmHg
(whichever is
greater)
(Mean × 0.86 +
35)mmHg
(Mean + 15) or
95mmHg
(whichever is
smaller)
Adult: 30 to 245
Pediatric: 30 to 180
Neonate: 25 to 105
IBP: PA
IBP-S SYS × 0.75 SYS × 0.75 SYS × 1.25 SYS × 1.25
3 to 120mmHg
IBP-D Dia × 0.75 Dia × 0.75 Dia × 1.25 Dia × 1.25
IBP-M Mean × 0.75 Mean × 0.75 Mean × 1.25 Mean × 1.25
IBP CPP
CPP × 0.68 +
8mmHg
(CPP – 15) or
35mmHg
(whichever is
greater)
CPP × 0.86+
35mmHg
(CPP+15) or
95mmHg
(whichever is
smaller)
Adult: 20 to 235 mmHg
Pediatric: 25 to 175 mmHg
Neonate: 25 to 100 mmHg