7-9
Module Parameter
Low alarm limit High alarm limit
Auto alarm limits range
Adult/
pediatric
Neonate
Adult/
pediatric
Neonate
IBP:
CVP/
ICP/
LAP/
RAP/
UVP/
P1-P4
(Venous
pressure)
IBP-M Mean × 0.75 Mean × 0.75 Mean × 1.25 Mean × 1.25 3 to 40mmHg
CO
2
EtCO
2
0 to
32mmHg:
remains the
same
0 to 32mmHg:
remains the
same
0 to 32mmHg:
remains the
same
0 to 32mmHg:
remains the same
Same as the measurement range
32 to
35mmHg:
29mmHg
32 to 35mmHg:
29mmHg
32 to 35mmHg:
41mmHg
32 to 35mmHg:
41mmHg
35 to
45mmHg:
(etCO
2
-6)
mmHg
35 to 45mmHg:
(etCO
2
-6) mmHg
35 to 45mmHg:
(etCO
2
+6)
mmHg
35 to 45mmHg:
(etCO
2
+6) mmHg
45 to
48mmHg:39
mmHg
45 to
48mmHg:39
mmHg
45 to
48mmHg:51
mmHg
45 to 48mmHg:51
mmHg
>48mmHg:
remains the
same
>48mmHg:
remains the
same
>48mmHg:
remains the
same
>48mmHg:
remains the same
FiCO
2
N/A N/A
Same as the
default alarm
limit
Same as the
default alarm limit
Same as the measurement range
awRR
awRR × 0.5 or
6 rpm
(whichever is
greater)
(awRR – 10) or 30
rpm (whichever
is greater)
awRR × 1.5 or
30 rpm
(whichever is
smaller)
(awRR+25) or 85
rpm (whichever is
smaller)
Adult/pediatric: 6 to 55
Neonate: 10 to 90