EasyManuals Logo

Mindray uMEC12 Operator's Manual

Mindray uMEC12
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #272 background imageLoading...
Page #272 background image
D-6
Measurement Alarm message L I Cause and solution
YY Non-Pulsatile L A The catheter may be occluded. Please flush
the catheter.
YY represents an IBP label.
C.O. TB Sensor Off L A Check the sensor connection and reconnect
the sensor.
CO
2
CO
2
Sensor High
Temp
L C Check, stop using or replace the sensor.
CO
2
Sensor Low
Temp
L C Check, stop using or replace the sensor.
CO
2
Temp Overrange L C The operating temperature of the CO2
module goes beyond the specified range.
After it restores within the specified range,
the module will restart automatically.
CO
2
High Airway
Press.
L C An error occurred in the airway pressure.
Check the patient connection and patient
circuit, and then restart the monitor.
CO
2
Low Airway
Press.
L C
CO
2
High Barometric L C Check the CO2 connections, make sure that
the monitor application site meets the
requirements, and check for special sources
that affect the ambient pressure. Restart the
monitor.
CO
2
Low Barometric L C
CO
2
FilterLine
Occluded
L
C
The airway or watertrap was occluded.
Check the airway and remove the occlusion.
CO
2
No Watertrap L B Check the watertrap connections.
CO
2
Check Adapter L A There is a problem with the airway adapter.
Check, clean or replace the adapter.
CO
2
FilterLine Err L C Check if there is a leak in the CO
2
sample
line or the CO
2
sample line has been
occluded.
CO
2
Zero Failed L A Check the CO
2
connections. After the
sensor’s temperature becomes stabilized,
perform a zero calibration again.
CO
2
System Err L A Re-plug the module or restart the monitor.
CO
2
Check Cal. L C Perform a calibration.
CO
2
Check Airway L C An error occurred to the airway.
CO
2
No Filterline L A Make sure that the filterline is connected.
CO
2
No Sensor L A Make sure that the sensor is connected.
CO
2
Main Board Err H C There is a problem with the CO
2
module.

Table of Contents

Other manuals for Mindray uMEC12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC12 and is the answer not in the manual?

Mindray uMEC12 Specifications

General IconGeneral
TypePatient Monitor
Display12.1-inch color TFT
ECG3-lead
SpO2Yes
NIBPYes
RespirationYes
TemperatureYes
BatteryRechargeable lithium-ion
NetworkingEthernet
ParametersECG, SpO2, NIBP, Respiration, Temperature
ConnectivityEthernet
Battery LifeUp to 4 hours
Power Supply100-240 VAC, 50/60 Hz

Related product manuals