D-7
Measurement Alarm message L I Cause and solution
CO
2
Checking Sensor L C
Re-plug the module or restart the monitor.
CO
2
Replace
Scrubber&Pump
L C
CO
2
15V Overrange H C
CO
2
Hardware Err H C
CO
2
Calibration
Required
L C
Perform a calibration.
CO
2
: Change
Watertrap
L C
Replace the watertrap.
CO
2
: Watertrap and
Patient Mismatch
L C
Check the patient category and use a
correct watertrap.
Power 12V Too High H C There is a problem with the system power
supply. Restart the monitor.
12V Too Low H C
5V Too High H C
5V Too Low H C
3.3V Too High H C
3.3V Too Low H C
Battery Too Low H C Connect the monitor to an AC power source
and allow the batteries to charge.
Different Battery
Voltages
M C The two batteries have different charge
capacity, or the batteries unspecified have
been used, or there is a problem with the
batteries. Make sure that correct batteries
are used and the batteries are not damaged,
or replace the batteries.
Battery Overload H C The power consumption of the equipment is
too high. Power the monitor using an AC
power source.
RT Clock Not Exist H C Contact your service personnel.
Recorder Recorder Init Err N L A Restart the monitor.
N is within 1 to 8.
Recorder SelfTest Err L A Stop the recording and restart the monitor.
Recorder Comm Err L A
Recorder S. Comm
Err
L A
Recorder Unavailable L A
Recorder Vlt High L C An error occurred to the system power