D-8
Measurement Alarm message L I Cause and solution
Recorder Vlt Low L C
supply. Restart the monitor.
Recorder Head Hot L C The recorder has been working for too long
time. Stop the recording and resume the
recording till the recorder’s printhead cools
down.
Rec Paper Wrong Pos. L A Re-load the recorder paper.
System System Watchdog Err H C An error occurred to the system. Restart the
monitor.
System Software Err H C
System CMOS Full H C
System CMOS Err H C
System FPGA Err H C
System Err N H C
N is within 2 to 12.
Other Bed
Disconnected
L A Check network connection.
PWR interrupted.
Check meas. state
L A Power supply failed accidently. Check the
measurements when the monitor restarts.
No CMS L A The monitor is disconnected from the CMS.
Check network connection.

Table of Contents

Other manuals for Mindray uMEC12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC12 and is the answer not in the manual?

Mindray uMEC12 Specifications

General IconGeneral
TypePatient Monitor
Display12.1-inch color TFT
ECG3-lead
SpO2Yes
NIBPYes
RespirationYes
TemperatureYes
BatteryRechargeable lithium-ion
NetworkingEthernet
ParametersECG, SpO2, NIBP, Respiration, Temperature
ConnectivityEthernet
Battery LifeUp to 4 hours
Power Supply100-240 VAC, 50/60 Hz

Related product manuals