Running an Experiment
188 0112-0109 H
Luminescence Assay Protocol
This section provides the following information and procedures:
• Materials required for the assay
• Cell preparation guidelines
• Coelentrazine loading procedure
• Running the assay on FLIPR
®
Tetra System
• Optimizing tips for Luminescence assays
Required Materials
The following materials are required for running a luminescence assay:
Table 8-3 Cell Preparation for Aequorin Assays.
Addition Parameters 96-Well Plate 384-Well Plate
Compound Addition Volume (μL) 50 25
Dispense Speed (μLs)
Adherent Cells
50–100 10–50
Dispense Speed (μL/s)
Non-adherent Cells
10–50 5–20
Note: The references to the instrument assume the Aequorin ICCD
camera option is present along with the Cell Suspension option.
Item Source
FLIPR
®
Tetra System with Aequorin,
ICCD camera option and Cell
Suspension option
Molecular Devices
Clear, flat-bottom, black wall 96-,
384-, 1536-well plates
See Appendix C
Clear polypropylene source plates (for
adherent protocols)
See Appendix C
Black FLIPR
®
System pipette tips, 96-
or 384-well or 1536 tip block and
gaskets
See Appendix C
Test compounds Specific to receptor
Cells in suspension Optional
Coelenterazine Specific to Photoproteins
Growth medium Major Laboratory Supplier (MLS)