Startup and Shutdown
42 0112-0109 H
Shutting Down the System
To shut down the FLIPR
®
Tetra System:
1. At the end of a programmed experiment, wait for the Assay
Finished (Unlocked) light on the Instrument Status Panel
to turn green, indicating the experiment is finished.
2. We recommend making sure that tips are removed from the
pipettor head if the last protocol did not remove them. You can
do this via manual command. Failure to remove tips can result
in an error on next start up.
3. Exit ScreenWorks Software by choosing Exit from the File
menu.
4. Turn off the computer and monitor.
5. Turn off the FLIPR
®
Tetra System power switch.
6. Turn off the chiller power switch.