Open Folder—An option which opens the Windows Explorer dialog box
directly to the directory designated.
Temperature Control
The Temperature Control field allows you to set the plate-stage
heaters to a desired temperature for an experiment. Temperature
range is from ambient 25 °C to 40 °C. Select Manual Settings to
control the heaters manually from ScreenWorks Software at the time
the experiment is run.
If, at the start of an experiment, the stage temperature is different
from the selected value a dialog informs you of this. You can wait until
the stage reaches the desired temperature or run the experiment
anyway.
Note: When the temperature is manually turned on through the
Instrument menu, a setting of “0” in the protocol does not affect the
temperature setting. If the user prefers to run off the temperature,
this must be manually done through Instrument > Manual
Operations > Temperature Control Off.