Robotic Integration
254 0112-0109 H
Command Parameters
There are no parameters associated with this command.
Example Command Line
stopexperiment<CR>
Example Response
ok<CR> or c55, error string <CR>
Suggested Steps in Command Usage
• Check if instrument is functioning fine.
• Make sure experiment is running.
• Send stopexperiment command.
• Response will be OK, confirming that command execution has
begun, or a Condition code related to the command indicating
that execution could not begin.
• Monitor status for completion, and error status for any other
kind of error.
Command Specific Error Responses
• Instrument is not functionalCondition code is 00
• No experiment is runningCondition code is 55
Clearerror<CR>
Command Description
Cleans error in case recoverable error occurred.
Command Parameters
There are no parameters associated with this command.
Example Command Line
clearerror<CR>
Example Response
ok<CR>
Suggested Steps in Command Usage
• Check instrument status.
• Check if error is occurred.
• Send clearerror command.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Molecular Devices FLIPR Tetra and is the answer not in the manual?

Molecular Devices FLIPR Tetra Specifications

General IconGeneral
BrandMolecular Devices
ModelFLIPR Tetra
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals