260 0112-0109 H
Example Command Line
washtips<TAB>a,2,28.0,20.00,fast,5,0.0,20.0<CR>
Example Response
ok<CR>
Suggested Steps in Command Usage
• Make sure an experiment is not running.
• Ensure that tips are loaded.
• Send washtips command.
• Response will be OK, confirming that command execution has
begun, or a Condition code related to the command indicating
that execution could not begin.
• Monitor status until the wash is complete.
Command Specific Error Responses
• Tips not loadedC68
Configuration <CR>
Command Description:
The configuration command returns the configure of the instrument.
The configuration response will be an ASCII text string which includes:
• Information about the instrument type and firmware version.
• Pipettor head’s type.
• Tip washer’s type.
• Bar code reader installed or not.
• Plate handler installed or not.
• Cell Reservoir installed or not.
Dispense
speed
Pipettor aspiration
speed in micro liters
per second.
5–200
(double)
1–50
(double)
.1–10
(double)
Table A-1 Washtips<TAB>Fluid Type, Wash Cycles, Volume/Stroke,
Aspirate Speed, Pump Speed, Strokes, Hold Time, Dispense
Speed<CR> command parameters (cont’d)
Parameter Description
Acceptable values or range
96-well
head
384-well
head
1536-
well head

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Molecular Devices FLIPR Tetra and is the answer not in the manual?

Molecular Devices FLIPR Tetra Specifications

General IconGeneral
BrandMolecular Devices
ModelFLIPR Tetra
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals