Robotic Integration
262 0112-0109 H
bar_code<SP>installed<TAB>
plate_handler<SP>installed<TAB>
cell_reservoir<SP>installed<TAB>
camera_installed<SP>emccd<TAB>
chiller_installed<SP>installed<TAB>
wave1<SP>470-495 nm<TAB>
wave2<SP>510-545 nm<TAB>
emfilt1<SP>515-575 nm<TAB>
emfilt2<SP>515-575 nm<TAB>
emfilt3<SP>515-575 nm<CR>
Suggested Steps in Command Usage
• Send configuration command.
• Response will be instrument configuration.
Statusex<CR>
Command Description
The statusex command is an extension of status command. It returns
the status of the instrument in more details than the status command.
The statusex response will be an ASCII text string which includes all
status command’s response information plus:
• Tip washer status
• Cell Reservoir status
• Camera chiller status
• Camera intensifier status
Example Command line
statusex<CR>
Response format
instrument <SP>instrument name<TAB>
version<SP>mm/dd/yyyy<TAB>
function<SP>ok|errorcode|fatalcode,[errorstring]<TAB>
inst_status<SP>busy|fault|error|resetting|abort|idle,remote|manual,o
ffline|online<TAB>
exp_running<SP>yes|no<TAB>
readplate<SP>platename|none,have|empty<TAB>
sourceplate1<SP>platename|none,have|empty<TAB>

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Molecular Devices FLIPR Tetra and is the answer not in the manual?

Molecular Devices FLIPR Tetra Specifications

General IconGeneral
BrandMolecular Devices
ModelFLIPR Tetra
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals