Loading...
HF group off
IMM PGM MIP
USAGE
SYNTAX xxHF or xxHF?
PARAMETERS
Description xx [ int ] - group number
Range xx - 1 to MAX GROUPS
Units xx - none
Defaults xx missing: error 13, GROUP NUMBER MISSING
out of range: error 14, GROUP NUMBER OUT OF RANGE
not assigned: error 15, GROUP NUMBER NOT ASSIGNED
floating point: truncated
DESCRIPTION This command turns power OFF of all axes assigned to a group. Refer MF
command to turn the power OFF of individual axes. The group power is
assumed to be OFF if power to any one of the axes in the group is OFF.
RETURNS If “?” sign is issued along with command, the controller returns:
1 - group power is ON
0 - group power is OFF
REL. COMMANDS HN - create a new group
HO - turn group power ON
EXAMPLE 1HN1,2 | create a new group (#1) with physical axes 1 and 2
1HO | turn group #1 power ON
1HF? | query group #1 power status
1 | controller returns a value of 1
1HF | turn group #1 power OFF
1HF? | query group #1 power status
0 | controller returns a value of 0
Section 3 – Remote Mode 3-71