Page #197 background imageLoading...
Page #197 background image
XX erase program
IMM PGM MIP
USAGE
SYNTAX xxXX
PARAMETERS
Description xx [ int ] - program number
Range xx - 1 to 100
Units xx - none
Defaults xx missing: error 38, COMMAND PARAMETER MISSING
out of range: error 7, PARAMETER OUT OF RANGE
DESCRIPTION This command makes the program xx loaded in controller’s non-volatile memory
unavailable to user. It does not delete the program from memory. Consequently,
the program space does not become available to user immediately after deleting
the program. It becomes available to user only after the entire stored program
memory is purged by issuing the command “0xx”.
Note:
Purging the stored program memory takes approximately 3 seconds for
completion.
RETURNS None
REL. COMMANDS EP - enter program download mode
EX - execute a stored program
LP - list stored program
XM - read available memory
EXAMPLE 1XX | delete program #1
XM | read available memory
Available storage space = 60228 | controller reports available storage space
2XX | delete program #2
XM | read available memory
Available storage space = 60228 | controller reports available storage space
0XX | purge stored program memory
XM | read available memory
Available storage space = 61440 | controller reports available storage space
Section 3 – Remote Mode 3-157

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Newport ESP300 Series and is the answer not in the manual?

Newport ESP300 Series Specifications

General IconGeneral
BrandNewport
ModelESP300 Series
CategoryController
LanguageEnglish

Related product manuals