Remote control commands
R&S
®
LCX Series
211User Manual 1179.2260.02 ─ 02
Usage: Query only
STATus:QUEStionable:NTRansition <Value>
Sets or queries the negative transition filter. If a bit is set, a 1 to 0 transition in the cor-
responding bit of the condition register causes a 1 to be written in the corresponding bit
of the event register.
Parameters:
<Value> Bit mask in decimal representation
Range: 1 to 65535
STATus:QUEStionable:PTRansition <Value>
Sets or queries the positive transition filter. If a bit is set, a 0 to 1 transition in the corre-
sponding bit of the condition register causes a 1 to be written in the corresponding bit
of the event register.
Parameters:
<Value> Bit mask in decimal representation
Range: 1 to 65535
STATus:QUEStionable[:EVENt]?
Queries the contents of the EVENt part of the status register to check whether an
event has occurred since the last reading. Reading an EVENt register deletes its con-
tents.
Return values:
<QUEStionable> Event bits in decimal representation
Range: 1 to 65535
Usage: Query only
11.15 SYSTem subsystem
The SYSTem subsystem contains the commands for general functions which do not
directly affect instrument operation.
SYSTem subsystem

Table of Contents

Other manuals for R&S LCX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S LCX Series and is the answer not in the manual?

R&S LCX Series Specifications

General IconGeneral
BrandR&S
ModelLCX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals