EasyManuals Logo

R&S LCX Series User Manual

R&S LCX Series
242 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #220 background imageLoading...
Page #220 background image
Remote control commands
R&S
®
LCX Series
220User Manual 1179.2260.02 ─ 02
SYSTem:HW:VERSion?
Queries the hardware version of the instrument.
Example: See Example "General instrument settings" on page 212.
Usage: Query only
Manual operation: See "Device Information" on page 129
SYSTem:KEY:BRIGhtness <Front key brightness>
Sets the brightness of the front panel keys.
Parameters:
<Front key
brightness>
Range: 0.1 to 1
Increment: 0.1
*RST: 0.8
Default unit: %
Example: See Example "General instrument settings" on page 212.
Manual operation: See "Key Brightness" on page 125
SYSTem:LOCal
Enables manual operation, i.e. unlocks front panel control.
To lock manual control, use command SYSTem:RWLock on page 221.
Example:
See Example "General instrument settings" on page 212.
Usage: Event
SYSTem:REMote
Activates remote control.
The R&S LCX switches to remote state, and locks all front panel controls.
You can control the R&S LCX remotely. Sending a command sets the instrument to
remote state, indicated by the white SCPI icon in the status bar.
To return to manual control, use command SYSTem:LOCal on page 220.
Example:
See Example "General instrument settings" on page 212.
Usage: Event
SYSTem:RESTart
Restarts the instrument without restarting the operating system.
Example:
See Example "General instrument settings" on page 212.
Usage: Event
SYSTem subsystem

Table of Contents

Other manuals for R&S LCX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S LCX Series and is the answer not in the manual?

R&S LCX Series Specifications

General IconGeneral
BrandR&S
ModelLCX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals