EasyManuals Logo

R&S LCX Series User Manual

R&S LCX Series
242 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #185 background imageLoading...
Page #185 background image
Remote control commands
R&S
®
LCX Series
185User Manual 1179.2260.02 ─ 02
CORRection:SPOT<Spot>:LOAD:STANdard <Reference value for primary>,
<Reference value for secondary>
Defines a working point for load correction.
Assign the working point number and the primary and secondary reference values.
Suffix:
<Spot>
.
1..n
Parameters:
<Reference value for
primary>
Sets the primary standard value as reference, e.g. the value of a
calibration resistor.
Default unit: Ohm
<Reference value for
secondary>
Sets the secondary standard value.
Default unit: Ohm
Example: See Example "Setting correction prameters" on page 181.
CORRection:SPOT<Spot>:LOAD[:EXECute]
Executes a load correction at a dedicated working point.
Suffix:
<Spot>
.
1..n
Example: See Example "Setting correction prameters" on page 181.
Usage: Event
Manual operation: See "Start Load Correction (Spot)" on page 80
CORRection:SPOT<Spot>:SHORt[:EXECute]
Executes a short correction at a dedicated working point.
Suffix:
<Spot>
.
1..n
Example: See Example "Setting correction prameters" on page 181.
Manual operation: See "Start Short/Open Correction (Spot)" on page 79
CORRection:SPOT<Spot>:STATe <Spot State>
Activates the specified working point for the correction function.
Suffix:
<Spot>
.
1..n
Parameters:
<Spot State> 1 | 0 | ON | OFF
*RST: 0|OFF
CORRection subsystem

Table of Contents

Other manuals for R&S LCX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S LCX Series and is the answer not in the manual?

R&S LCX Series Specifications

General IconGeneral
BrandR&S
ModelLCX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals