Remote control commands
R&S
®
LCX Series
206User Manual 1179.2260.02 ─ 02
Example: Acquiring measurement readings
The example represents the remote control commands of the measurement results.
**************************************************
// Request information on the measurement result
// query recently measured impedance values
READ:IMPedance?
// Response: returns Z and Theta d values
// query the recently measured values of measurement pair 2
READ?
// Response: returns the measurement values of he selected parameter pair
// Request information on the measurement result
// query recently measured values of the impedance
// Request most recent valid readings
// Query the valid impedance measurement values
FETCh:IMPedance?
// Response: returns Z and Theata d
// Query the valid values of measurement pair 2
FETCh?
// Response: returns the valid values of the selected parameter pair
FETCh?........................................................................................................................ 206
FETCh:IMPedance?.......................................................................................................206
INITiate[:IMMediate]....................................................................................................... 207
MEASure:ACCuracy?.....................................................................................................207
MEASure:CURRent?......................................................................................................207
MEASure:MODE............................................................................................................207
MEASure:TRIGger:DELay.............................................................................................. 208
MEASure:VOLTage?...................................................................................................... 208
READ?......................................................................................................................... 208
READ:IMPedance?........................................................................................................ 208
FETCh?
Queries the most recent valid values for measurement pair 2.
If no valid measurement values are available, the reponse reports error code -230.
Example:
See Example "Acquiring measurement readings" on page 206.
Usage: Query only
FETCh:IMPedance?
Queries the most recent valid values of the measured impedance.
If no valid measurement values are available, the reponse reports error code -230.
Example:
See Example "Acquiring measurement readings" on page 206.