Remote control commands
R&S
®
LCX Series
198User Manual 1179.2260.02 ─ 02
FUNCtion:MEASurement:TYPE <Measurement Type>
FUNCtion:MEASurement:TYPE? <Measurement Type>
Selects the measurement function.
Parameters for setting and query:
<Measurement Type> L | C | R | T
L
Impedance measurement.
C
Capacitance measurement.
R
Resistance measurement.
T
Transformer measurement.
Example: See Example "Selecting the measurement function and setting
the impedance range" on page 195.
Usage: Asynchronous command
Manual operation: See "Functions [hardkeys]" on page 70
FUNCtion:TRANsformer:RANGe[:TYPE] <Turn Ratio>
FUNCtion:TRANsformer:RANGe[:TYPE]? <Turn Ratio>
Selects the impedance range for transformer measurement.
Parameters for setting and query:
<Turn Ratio> N50 | N500
Example: See Example "Selecting the measurement function and setting
the impedance range" on page 195.
11.11 HANDler subsystem
The HANDler subsystem contains the commands for configuring the binning function.
HANDler subsystem